Clinical Solutions
Cirrus™ HD-OCT – Third Party Hardware and Software
The following hardware and software is compatible with the Cirrus™ HD-OCT.

Software Version 5
The Cirrus Version 5 Software has been tested with McAfee Virus Scan Enterprise (version 8.5.0 i), Kaspersky Anti-Virus (version 7.0.0.125) and Microsoft® Security Essentials (version 1.0.1611.0), but anti-virus software does not come pre-installed.
When installing your anti-virus software per manufacturer’s recommendations, configure it so it does not scan automatically, as this may interrupt the operation of the Cirrus instrument or your review station.

The Cirrus Version 5 Software is compatible with Windows XP Service Packs 2 and 3, and with Windows Vista Service Packs 1 and 2.


Compatible Printers
  • HP 6980
  • Brother HL-2700CN
  • HP 6988
  • HP 6000

Restrictions (see notes): 4, 5, 6, 7 & 8.


Network Switches

Generic
  • Interface speed:
    10/100/1000 Mb/s
  • Interface connectors:
    RJ-45
  • Cable detection:
    MDI/MDI-X (automatic straight-through or crossover cable detection)
  • Supported standards:
    IEEE 802.3, 802.3u, 802.3ab, and 802.3x
  • Device must have CE approvals and FCC approvals.


Network Attached Storage
  • NAS Drive
    Restrictions (see notes): 4, 5, 6 & 7.


Other External Devices
  • External DVD-RAM Drive
    Restrictions (see notes): 5 & 7
  • USB Hub
    Restrictions (see notes): 1, 2 & 3.
  • USB (Key Type) Storage Device
    Restrictions (see notes): 1, 2 & 3.

Restriction notes
  1. Device must have CE approvals.
  2. Device must have FCC approvals.
  3. Device must be powered from USB port only.
  4. Device must be outside the patient environment (at least 1.5 meters away), unless used with an isolation transformer.
  5. Do not use with an extension cord or power strip.
  6. Use unshielded network cables (UTP) only.
  7. Do not share the same wall outlet with the Cirrus HD-OCT.
  8. Contact Carl Zeiss Meditec for compatible printer
    driver version(s).
More information